The encyclopedia · Engineering & Operations · Technical decision · 2025
Tandem's t:slim X2 speaker wiring can stop insulin delivery mid-use
A speaker wiring defect triggers a Malfunction 16 alarm that halts insulin delivery and cuts CGM communication; 700 adverse events, 59 injuries before the fix.
Tandem Diabetes Care · 2025-08-07
What happened
In July and August 2025, Tandem Diabetes Care issued a voluntary medical device correction for select t:slim X2 insulin pumps. A wiring issue with certain pumps' speakers could trigger an error presenting as a 'Malfunction 16' alarm that stops insulin delivery and terminates communication between the pump, the continuous glucose monitor and the companion app.
The consequence was loss of insulin and of live glucose data, with a risk of hyperglycemia severe enough that some patients required medical intervention. By the time of the announcement Tandem had tallied 700 confirmed adverse events, 59 reported injuries and no deaths.
Tandem said it notified customers starting 22 July 2025, would release a software update for early detection of speaker failure and persistent vibration alerts, and asked users to complete the update when notified. The FDA posted the correction on 12 August 2025.
The decision under review was a component's failure mode reaching production: a cheap speaker whose wiring could fail in a device patients rely on to stay alive, caught only after hundreds of adverse events rather than in design review.
Why it happened
- A low-cost component carried a failure mode that stopped a life-sustaining device
- The defect surfaced only after 700 adverse events and 59 injuries, not in validation
- The fix was a software update for detection, leaving the physical wiring risk in place
- Trust in a device that keeps people alive was the real cost, on top of the recall
The lesson
In a device that keeps someone alive, a cheap component's failure mode must be caught before it ships, not patched after.
Sources
- Tandem Diabetes Care issues voluntary medical device correction for select t:slim X2 insulin pumps — FDA, August 2025
- Tandem's t:slim insulin pump recall — MedTech Dive, August 2025
spotted an error? The club wants to know.
More like this
Union Carbide's Bhopal plant leaked toxic gas — thousands died in their sleep
Optiray 350: glass and steel in the contrast vial
Adequan: glass fiber in the joint injection for dogs and horses
Somewhere, someone solved the problem this company failed at. 2nd Opinion →

Comments · 0
Sign in to join the comments.